Excel Precision1510A sensor is designed for measurement that requires high resolution and frequency response.   The complete system consists of one each of 1511A Laser head, 1521A Sensor and 1522A Signal Processor.   The 1511 Laser Head and 1521A Sensor are independent units and can be mounted individually, which provides excellent measurement flexibility. spacer.gif (807 bytes) When there is a relative movement between the target and the reference object, the beam will shine on different locations of the detector.   The user is able to determine the amount of movement comparing the signal intensity on the detector.   Multiple measurement ranges and resolutions are available for best measurement results.

spacer.gif (807 bytes) Specifications:
Dimensions:   Laser Head    0.75"(H) x 0.59"(W) x 1.98"(L) (1511)
                       Detector          1.43" diameter (1521)
Light Source:              Semiconductor Laser (652nm, 5mW)
Beam Size:                 2mm x 4mm collimated 
Reference Distance:    Up to 350 mm
Working Range:         1mm in lateral direction
Resolution:                 Max. 0.1 micron @ 1 KHz
Sensitivity:                  Max. 0.1 micron / 10 mV
Repeatability:             1%
Frequency Response: Up to 1KHz
System Power Requirement:  +/- 15V +/- 5%
Signal Output:  Analog, -10V to +10V           
spacer.gif (807 bytes)